Published 2011 | Version v1
Journal article

Evaluation of radiation dose during pacemaker implantations

  • 1. Dept.s of Cardiology and Radiology, Konstantopoulio General Hospital, 3-5 Agias Olgas Atr, 14233 Nea Ionia, Athens (Greece)

Description

The purpose was to evaluate patient radiation doses and compare with other interventional procedures. One hundred and twenty-eight procedures were carried out with a recently installed mobile under-couch C-arm machine with a 23-cm diameter image intensifier. The radiation dose is provided by the X-ray machine, in terms of cumulative dose (CD). Kerma-Area product (KAP) was then estimated from CD and the X-ray field size. Other patient parameters recorded were patient weight, age, kilovolt, milli ampere and fluoroscopy time (T). Median (range) CD, KAP and T were 15.2 mGy (3.2-110 mGy), 6.3 Gy cm2 (1.3-45.7 Gy cm2) and 5.2 (1.5-27.4 min) min, respectively. Median E was 1.1 mSv (conversion factor: 0.18 mSv per Gy cm2), which corresponds to approximately one lumbar spine X-ray radiography. The effective dose is much lower than a coronary angiography (8 mSv) or an electrophysiology study (6 mSv). Radiation dose is low compared with other interventional cardiology procedures. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncr267

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
147
Journal Issue
1-2
Journal Page Range
p. 75-77
ISSN
0144-8420

Conference

Title
International Conference on Radiation Protection in Medicine
Dates
1-3 Sep 2010
Place
Varna (Bulgaria)

Optional Information

Notes
6 refs