Optimizing image quality and dose for digital radiography of distal pediatric extremities using the contrast-to-noise ratio

  • 1. University of Applied Science, Hamburg (Germany)
  • 2. Philips Healthcare, Hamburg (Germany). Clinical Science Diagnostic X-Ray

Description

Purpose: To investigate the influence of X-ray tube voltage and filtration on image quality in terms of contrast-to-noise ratio (CNR) and dose for digital radiography of distal pediatric extremities and to determine conditions that give the best balance of CNR and patient dose. Materials and Methods: In a phantom study simulating the absorption properties of distal extremities, the CNR and the related patient dose were determined as a function of tube voltage in the range 40 - 66 kV, both with and without additional filtration of 0.1 mm Cu/1 mm Al. The measured CNR was used as an indicator of image quality, while the mean absorbed dose (MAD) - determined by a combination of measurement and simulation - was used as an indicator of the patient dose. Results: The most favorable relation of CNR and dose was found for the lowest tube voltage investigated (40 kV) without additional filtration. Compared to a situation with 50 kV or 60 kV, the mean absorbed dose could be lowered by 24 % and 50 %, respectively, while keeping the image quality (CNR) at the same level. Conclusion: For digital radiography of distal pediatric extremities, further CNR and dose optimization appears to be possible using lower tube voltages. Further clinical investigation of the suggested parameters is necessary. (orig.)

Additional details

Publishing Information

Journal Title
RoeFo - Fortschritte auf dem Gebiete der Roentgenstrahlen und der bildgebenden Verfahren
Journal Volume
184
Journal Issue
7
Journal Page Range
p. 643-649
ISSN
1438-9029
CODEN
RFGNDO