Published January 1990 | Version v1
Journal article

Evaluation of proton density by magnetic resonance imaging: phantom experiments and analysis of multiple component proton transverse relaxation

  • 1. German Cancer Research Center, Heidelberg (Germany, F.R.). Inst. of Radiology and Pathophysiology

Description

This study was undertaken to estimate the accuracy of proton density measurements using a standard whole-body MR imager operating at 1.5 T. It was concluded that the hydrogen density of biological tissues can be evaluated at best with an overall error of 10% from MR images for TE>10 ms. This accuracy is insufficient for a pixel-orientated neutron therapy planning. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
35
Journal Issue
1
Series
Phys. Med. Biol.
Journal Page Range
53-66
ISSN
0031-9155
CODEN
PHMBA

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
21040032
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; CORRECTIONS; ERRORS; HYDROGEN; MAGNETIC RESONANCE; NEUTRON THERAPY; PHANTOMS; PLANNING; PROTON DENSITY; TISSUES
Descriptors DEC
BODY; ELEMENTS; MEDICINE; MOCKUP; NONMETALS; RADIOTHERAPY; RESONANCE; STRUCTURAL MODELS; THERAPY