Published January 1990
| Version v1
Journal article
Evaluation of proton density by magnetic resonance imaging: phantom experiments and analysis of multiple component proton transverse relaxation
Creators
- 1. German Cancer Research Center, Heidelberg (Germany, F.R.). Inst. of Radiology and Pathophysiology
Description
This study was undertaken to estimate the accuracy of proton density measurements using a standard whole-body MR imager operating at 1.5 T. It was concluded that the hydrogen density of biological tissues can be evaluated at best with an overall error of 10% from MR images for TE>10 ms. This accuracy is insufficient for a pixel-orientated neutron therapy planning. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 35
- Journal Issue
- 1
- Series
- Phys. Med. Biol.
- Journal Page Range
- 53-66
- ISSN
- 0031-9155
- CODEN
- PHMBA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 21040032
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; CORRECTIONS; ERRORS; HYDROGEN; MAGNETIC RESONANCE; NEUTRON THERAPY; PHANTOMS; PLANNING; PROTON DENSITY; TISSUES
- Descriptors DEC
- BODY; ELEMENTS; MEDICINE; MOCKUP; NONMETALS; RADIOTHERAPY; RESONANCE; STRUCTURAL MODELS; THERAPY