Improving CdZnTe X-ray detector performance by cooling and rise time discrimination

  • 1. Metorex Int. Oy, Espoo (Finland)
  • 2. Baltic Scientific Instruments Ltd., 26 Ganibu Dambis, LV-1005 Riga (Latvia)

Description

We have studied several methods for improving CdZnTe detector performance, especially energy resolution and charge collection properties. Energy resolution can be greatly improved by cooling the detector to about -30 C which reduces the detector leakage current enabling the use of low-noise, pulsed-optical feedback preamplifiers and higher bias voltages. High bias voltages are necessary for more efficient charge collection which reduces spectrum background and peak tailing. Applying rise time discrimination circuitry to the linear amplifier reduces the tailing effect even further, especially at higher energies. Combining all these methods we tested a 16-mm2, 3-mm thick Cd0.9Zn0.1Te detector crystal at -30 C, and obtained energy resolutions of 292 eV at the 5.9-keV line of 55Fe, 600 eV at the 59.5-keV line of 241Am, and 780 eV at the 88-keV line of 109Cd. Another 9-mm2, 3-mm thick Cd0.9Zn0.1Te detector crystal showed a charge collection problem at -30 C which caused all spectrum peaks to be doublets. However, the rise time discriminator removed the ghost peaks form the spectrum almost completely, and improved the energy resolution and background. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
377
Journal Issue
2-3
Journal Page Range
p. 484-486.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
7. European symposium on semiconductor detectors, new developments in radiation detectors.
Dates
7-10 May 1995.
Place
Schloss Elmau (Germany).