Published October 21, 2007 | Version v1
Journal article

Towards a PMT based optical readout GEM TPC-First results

  • 1. ISEC, Rua Pedro Nunes, Quinta da Nora, 3030-199 Coimbra (Portugal)
  • 2. LIP - Coimbra and Departamento de Fisica da Universidade de Coimbra, 3004-516 Coimbra (Portugal)
  • 3. CFRMUC, Departamento de Fisica da Universidade de Coimbra, 3004-516 Coimbra (Portugal)

Description

We report the first results on the operation of an optical double GEM TPC, read out by four photomultipliers in a square array. The detector, operated with Ar+5%CF4 at atmospheric pressure, was tested with 241Am alpha particles emitted in various orientations. We obtained a typical track angle resolution of 2 deg. (σ) at 45 deg. and achieved a point position resolution of better than 1 mm (FWHM) for the X- and Y-directions in the GEM plane

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2007.07.078

Additional details

Identifiers

DOI
10.1016/j.nima.2007.07.078;
PII
S0168-9002(07)01532-X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
581
Journal Issue
1-2
Journal Page Range
p. 202-205
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
11. international Vienna conference on instrumentation
Acronym
VCI 2007
Dates
19-21 Feb 2007
Place
Vienna (Austria)

Optional Information