Published September 2003 | Version v1
Journal article

k0-NAA quality assessment by analysis of different certified reference materials using the KAYZERO/SOLCOI software

  • 1. Jozef Stefan Institute, Ljubljana (Slovenia). Dept. of Environmental Sciences
  • 2. International Atomic Energy Agency, Vienna (Austria). Div. of Human Health

Description

A suite of natural matrix reference materials (RMs) were used to assess the quality of analytical results obtained by k0-instrumental neutron activation analysis (k0-INAA) at the Jozef Stefan Institute (IJS). Five certified reference materials (CRMs) from the Institute for Reference Materials and Measurements (IRMM), two standard reference materials (SRMs) from the National Institute of Standards and Technology (NIST), three RMs from the International Atomic Energy Agency (IAEA) and one RM from IJS were analyzed. Altogether, results for twenty-four elements in inorganic matrices and twenty-nine elements in organic matrices, obtained by k0-INAA, were compared to certified values. Results obtained show good agreement with certified or assigned values except for Fe, La, Nd, Sm and U in inorganic matrices, and Ag, Al and Cr in organic matrices. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
257
Journal Issue
3
Journal Page Range
p. 659-663
ISSN
0236-5731
CODEN
JRNCDM

Conference

Title
3. International k_0-users Workshop
Dates
23-28 Sep 2001
Place
Bruges (Belgium)

Optional Information

Notes
15 refs.