Metrology of the maximum temperature measurement method by means of IMIK

Description

Results of state testing of the crystalline maximum temperature meter (CMTM) have been presented. The root-mean-square error of a separate measurement under stationary and nonstationary conditions constitutes 6 and 12 deg, respectively. Comparison of statistical data with the data obtained by means of the CMTM permitted to obtain measurement equations excluding a part of the error systematic component. Dependences of the root-mean-square deviation (RMSD) on the measured temperature level are obtained both for the diamond and silicon carbide CMTM. RMSD dependence on the temperature has been also estimated by the calculation method based on common considerations. The minimum RMSD value (about 4-5 deg C) is found to be achieved in the 500-1000 deg C temperature range. Brief description of other methods of temperature measurement in a complicated areas for measurements of technical installations is presented and RMSD dependences on measured temperature are considered

Availability note (English)

MF available from INIS under the Report Number.

Files

12582428.pdf

Files (713.5 kB)

Name Size Download all
md5:1969c91d990fda22f98ce8a83a6391db
713.5 kB Preview Download

System files (71.4 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Метрология метода измерения максимальной температуры с помощью IMTK

Publishing Information

Imprint Pagination
25 p.
Report number
IAE--3208

Optional Information

Notes
15 refs.; 3 figs.; 8 tables.