Report on radiation exposure of lead-scintillator stack
Description
A stack of lead and scintillator was placed in a neutral beam obtained from targeting 800 GeV protons. Small pieces of film containing radiochromic dye were placed adjacent to the layers of scintillator for the purpose of measuring the radiation dose to the scintillator. Our motivation was to calibrate the radiation dose obtainable in this manner for future tests of scintillator for SSC experiments and to relate dose to flux to check absolute normalization for calculations. We also observed several other radiation effects which should be considered for both damage and compensation in a calorimeter
Availability note (English)
MF available from INIS under the Report Number; Available from OSTI as DE93040813; NTIS; US Govt. Printing Office Dep.Files
25019000.pdf
Files
(144.1 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:b42a7b6665e12ca34696a7a210c5be86
|
144.1 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 9 p.
- Report number
- ANL-HEP-TR--90-108
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 25019000
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S43: PARTICLE ACCELERATORS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; LEAD; PHYSICAL RADIATION EFFECTS; RADIATION DOSES; SOLID SCINTILLATION DETECTORS; SUPERCONDUCTING SUPER COLLIDER
- Descriptors DEC
- ACCELERATORS; CYCLIC ACCELERATORS; ELEMENTS; MEASURING INSTRUMENTS; METALS; RADIATION DETECTORS; RADIATION EFFECTS; SCINTILLATION COUNTERS; STORAGE RINGS; SYNCHROTRONS