The in-situ calibration method for ambient dose equivalent dosemeters
Creators
- 1. Radiation Metrology Center of China Institute of Atomic Energy 102413, -CIAE-, P.O. Box 275 -20- No. 1, Sanqiang Road, Fangshan District, Beijing (China)
Description
Ambient dose equivalent dosemeters are widely distributed in nuclear facilities for routine continuous monitoring or nuclear accident emergency monitoring, they are inconvenient to be disassembled and sent to the metrology laboratory for calibration. To ensure the accuracy of such dosemeters, the research of in-situ calibration method has been carried out. Ambient dose equivalent secondary standard ionization chamber and portable gamma ray irradiation facility have been designed for in-situ calibration. The experiments of in-situ calibration were carried out in five typical sites located in China Institute of Atomic Energy, results showed that the relative deviation between in-situ calibration factors and metrology laboratory calibration factors are within 5%.The combined standard uncertainty of the in-situ calibration factor is 2.8%, and the expanded uncertainty is 5.6% (k=2).The in-situ calibration method can meet the calibration requirements of the fixed ambient dose equivalent dosemeters. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncab130Additional details
Identifiers
- DOI
- 10.1093/rpd/ncab130;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 196
- Journal Issue
- 1-2
- Journal Page Range
- p. 85-94
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 53001942
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; AMBIENT DOSE EQUIVALENTS; CALIBRATION; DESIGN; DOSEMETERS; GAMMA RADIATION; IONIZATION CHAMBERS; IRRADIATION PLANTS; METROLOGY; RADIATION ACCIDENTS; RADIATION MONITORING; REACTOR ACCIDENTS; SAFETY STANDARDS
- Descriptors DEC
- ACCIDENTS; DOSE EQUIVALENTS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MONITORING; NUCLEAR FACILITIES; RADIATION DETECTORS; RADIATIONS; STANDARDS
Optional Information
- Notes
- 8 refs.