Published May 30, 1994 | Version v1
Journal article

Preconcentration of trace elements from high-purity thorium and uranium on Chelex-100 and determination by graphite furnace atomic absorption spectrometry with Zeeman-effect background correction

  • 1. Analytical Chemistry Division, BARC, Trombay, Bombay (India)

Description

Preconcentration of trace impurities from large-sized samples of uranium metal and thorium oxide using a small column of Chelex-100 followed by their determination using graphite furnace atomic absorption spectrometry (GFAAS) is reported. A 0.5-10-g amount of the sample (uranium metal or thorium oxide) was dissolved, complexed with ammonium carbonate and subjected to the ion-exchange procedure. The retained analytes were eluted with 2-4 M nitric acid and brought to a small volume for a final dilution to 10-25 ml for their determination using GFAAS. The validity of the separation procedure and recoveries at μg kg-1 levels was checked by standard addition; the recoveries were >95%

Additional details

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
290
Journal Issue
3
Journal Page Range
p. 371-377.
ISSN
0003-2670
CODEN
ACACAM