Published 1977
| Version v1
Book
Response of a commercial TL-Dosimeter to 9 MeV photon radiation
Creators
- 1. Instituto de Radioprotecao e Dosimetria, Rio de Janeiro (Brazil)
- 2. Comission of the European Communities, Joint Research Centre at Ispra, Italy
Description
It is shown that dosimeter earlier calibrated with Gamma-Radiation from Co-60 underestimates the absorbed dose due to 9 MeV-Gamma-Radiation in its maximum in water and/or tissue by a factor of 1.89 or nearly 2, if the chip behine the 10mg/cm2 filter is used and a factor of 1.38 if the chip behind the 285 mg/cm2 filter is evaluated
Additional details
Publishing Information
- Publisher
- A. Scharmann.
- Imprint Place
- Sao Paulo
- Imprint Title
- Proceedings of the Fifth International Conference on Luminescence Dosimetry
- Imprint Pagination
- 497 p.
- Journal Page Range
- p. 346-349.
Conference
- Title
- 5. International Conference on Luminescence Dosimetry.
- Dates
- 14 - 17 Feb 1977.
- Place
- Sao Paulo, Brazil.
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 10427830
- Subject category
- S74: ATOMIC AND MOLECULAR PHYSICS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; GAMMA RADIATION; MEV RANGE 01-10; PHOTONS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- DOSEMETERS; ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; ENERGY RANGE; IONIZING RADIATIONS; LUMINESCENT DOSEMETERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEV RANGE; RADIATIONS