Description

Skin toxicity has been reported for IMRT of head and neck cancer. The purpose of this study was to investigate the dose in the build-up region delivered by a 6 MV treatment plan for which important skin toxicity was observed. We also investigated if the different designs of the treatment head of an Elekta and a Varian linear accelerator, especially the lower position of the Varian multi-leaf collimator, give rise to different build-up doses. For regular square open beams, the build-up dose along the central beam axis is higher for the Varian machine than for the Elekta machine, both for 6 MV and 18 MV. At the Elekta machine at 18 MV, the superficial dose of a diamond shaped 10 x 10 cm2 field is 3.6% lower than the superficial dose of a regular 10 x 10 cm2 field. This effect is not seen at 6 MV. At the Varian machine, the superficial dose of the diamond shaped field is respectively 3.5 and 14.2% higher than the superficial dose of the regular 10 x 10 cm2 field for 6 MV and 18 MV. Despite the differences measured in build-up dose for single beams between the Elekta and the Varian linear accelerator, there were no measurable differences in superficial dose when a typical IMRT dose plan of 6 MV for a head and neck tumour is executed at the two machines

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/50/413/pmb5_3_002.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
50
Journal Issue
3
Journal Page Range
p. 413-428
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
36033612
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CARCINOMAS; COLLIMATORS; DESIGN; DIAMONDS; HEAD; LINEAR ACCELERATORS; NECK; PHOTON BEAMS; RADIATION DOSES; RADIOTHERAPY; SKIN; TOXICITY
Descriptors DEC
ACCELERATORS; BEAMS; BODY; CARBON; DISEASES; DOSES; ELEMENTS; MEDICINE; MINERALS; NEOPLASMS; NONMETALS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY