Published June 1989 | Version v1
Journal article

Physical aspects of quality assurance in cobalt teletherapy units

  • 1. Malaya Univ., Kuala Lumpur (Malaysia)

Description

ICRU Report No. 24 states that +- 5% is a reasonable degree of accuracy to strive for in the delivery of absorbed doses to target volumes in radiotherapy. To achieve this quality assurance level, a comprehensive programme comprising both physical and clinical components is needed. The purpose of this paper is to present the rationale and guidelines for implementing a recommended quality assurance programme. It will emphasize the physical aspects pertaining to acceptance and constancy tests using Cobalt-60 teletherapy units. It will also outline a suggested schedule of quality assurance testings including radiation dosimetry, geometry and electrochemical safety. (author)

Additional details

Publishing Information

Journal Title
Jurnal Sains Nuklear Malaysia
Journal Volume
7
Journal Issue
1
Series
J. Sains Nukl. Malays.
Journal Page Range
41-48
ISSN
0127-2810
CODEN
JSNME