Published October 1993 | Version v1
Journal article

Correction factors for photo peak efficiencies of a LEPS and 60ml Ge detectors due to source sizes

  • 1. Kinki Univ., Higashi-Osaka, Osaka (Japan). Faculty of Science and Technology

Description

The correction factors are obtained from the following two methods; (1) the relative distribution of the photo peak efficiencies of the detectors as a function of the position of a point source, which is measured by removing the source on the plane parallel with the surface of the detector at the plane-detector distances specified; and then the factors are calculated by intergating the efficiencies distributing within the circular regions of any of sizes; and (2) also obtained from the practical relation between the efficiencies, which is measured directly for circular disk sources and those for point sources. The factors thus obtained from the different methods are compared in those cases which are within a 15 mm radius. (author)

Additional details

Publishing Information

Journal Title
Kinki Daigaku Rikogakubu Kenkyu Hokoku
Journal Issue
no.29
Journal Page Range
p. 143-150.
ISSN
0386-4928
CODEN
KDRKBB