The use of total evaporation method using Channeltron electron multipliers by thermal ionization mass spectrometry for europium isotope ratio measurements on pico-gram sample amounts

  • 1. Commissariat Energie Atom, DEN DPC SECR LANIE, F-91191 Gif Sur Yvette, (France)
  • 2. Commissariat Energie Atom, DEN DPC, F-91191 Gif Sur Yvette, (France)

Description

Thermal ionization mass spectrometry (TIMS) is a well established instrumental technique for providing accurate and precise isotope ratio measurements for elements with low first ionization potential. In the nuclear domain, particularly for the study of spent nuclear fuel sample solutions, the reduction of quantities is an important issue in order to decrease the analyst exposition to the radioactive samples. This paper presents experimental results obtained for pico-gram quantities of samples using electron multipliers (Channeltrons) implanted on the new generation of TIMS associated with the total evaporation (TE) acquisition method. TE method presents some advantages for the measurement of low quantities of samples in relation to the whole sample consumption and the integration of the whole signal. For this study. a Thermal Ionization Mass Spectrometer from IsotopX (Isoprobe-T) equipped with a multi-collector system including Faraday cups and four movable Channeltrons is used. This work is focused on lanthanide elements, and particularly europium, in relation to the importance of their characterization in the spent fuel to validate neutronic calculation codes. In a first approach, europium isotopic ratios by total evaporation method using Faraday cups are obtained for quantities down to 250 pg. The results obtained by TE (Faraday cups only) on such low quantities confirm the potentialities of the method. A reproducibility of 2.5 parts per thousand (k = 2), was obtained on the 153Eu/151Eu ratio (n = 11), for natural samples with an accuracy around 1 parts per thousand. In a second approach the same procedure is tested on Channeltron detectors by reducing the amounts of samples (from 50 pg to 1 pg). In order to obtain accurate isotope results, intrinsic parameters such as applied voltages, dead time and noise were determined. A Channeltron specific intercalibration gain method was also developed, as this parameter was found to be the major source of uncertainties using this type of detectors. The reproducibility on the 153Eu/151Eu ratio was respectively found around 3% (k = 2). These results show the potentialities of isotopic measurements of sub-nanograms quantities and the different sources of uncertainties using total evaporation and Channeltrons on the lsoprobe-T. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1016/j.ijms.2011.09.014

Additional details

Identifiers

Publishing Information

Journal Title
International Journal of Mass Spectrometry (Print)
Journal Volume
309
Series
Country input: France
Journal Page Range
p. 141-147
ISSN
1387-3806

Optional Information

Notes
48 refs.