Published October 1988
| Version v1
Journal article
Energy calibration of organic scintillation detectors for γ rays
- 1. Institute of Nuclear Research, Academia Sinica, Shanghai
Description
An experimental method of calibrating organic detectors is described. A NaI(T1) detector has some advantages of high detection efficiency, good energy resolution, and definite position of the back-scattering peak. The precise position of the Compton edge can be determined by coincidence measurement between the pulse of an organic scintillation detector and the pulse of the back-scattering peak from NaI(T1) detector. It can be used to calibrate various sizes and shapes of organic scintillation detectors simply and reliably. The home-made plastic and organic liquid scintillation detectors are calibrated and positions of the Compton edge as a function of γ-ray energies are obtained
Additional details
Publishing Information
- Journal Title
- Chinese Physics
- Journal Volume
- 8
- Journal Issue
- 4
- Series
- Chin. Phys.
- Journal Page Range
- 1030-1034
- ISSN
- 0273-429X
- CODEN
- CHPHD
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20016672
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; ENERGY RESOLUTION; GAMMA DETECTION; LIQUID SCINTILLATION DETECTORS; NAI DETECTORS; ORGANIC COMPOUNDS; USES
- Descriptors DEC
- DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS