Dose-rate dependence of PAGAT polymer gel dosimeter evaluated using X-ray CT scanner
Description
In recent years, X-ray CT imaging of polymer gel dosimeter has been applied to direct measurement of 3D dose distributions. This work presents the dependence of PAGAT polymer gel dosimeter response to different beam dose rate using X-ray CT scanner as an evaluation tool. The normoxic PAGAT polymer gel was prepared on bench top and irradiated to different doses at different dose-rate ranging from 0.5 to 5 Gy/min. After irradiation the dosimeter samples were imaged with X-ray CT scanner using appropriate scanning protocol. The results showed that PAGAT gel dosimeter formulations investigated in this study has the smallest dose rate dependence and the change in dose sensitivity amounts to 2 % over the dose rate interval studied.
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/250/1/012074Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 250
- Journal Issue
- 1
- Journal Page Range
- [5 p.]
- ISSN
- 1742-6596
Conference
- Title
- 6. international conference on 3D radiation dosimetry
- Acronym
- IC3DDose
- Dates
- 22-26 Aug 2010
- Place
- Hilton Head Island, SC (United States)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42053589
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BEAMS; COMPUTERIZED TOMOGRAPHY; DOSE RATES; DOSEMETERS; GELS; IRRADIATION; POLYMERS; RADIATION DOSE DISTRIBUTIONS; SENSITIVITY; X RADIATION
- Descriptors DEC
- COLLOIDS; DIAGNOSTIC TECHNIQUES; DISPERSIONS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATIONS; TOMOGRAPHY