Applicability of a practical calibration for the small-type OSL dosimeter for measuring doses from direct X-rays and penetrating X-rays affected by scattered radiation
- 1. Nagase Landauer, Ltd., Tsukuba, Ibaraki (Japan)
- 2. Kanazawa University, Kanazawa, Ishikawa (Japan)
- 3. Tokushima University, Graduated School of Health Sciences, Tokushima (Japan)
Description
A small type optically stimulated luminescence dosimeter, nanoDot, is expected to be applied to radiation dose evaluation on medical field. Dose calibration of a nanoDot is carried out by the air Kerma from incident X-ray generated by a diagnostic X-ray equipment. However the main component of exposure is not only direct X-ray beam but also scattered X-rays and penetrating X-rays. Therefore we evaluated the reliability of the dose calibration of nanoDot dosimeters on the exposure of different quality X-rays. The main calibration curve was determined by the air Kerma of 83 kV X-rays in the dose range of 3 μGy to 2 mGy. To validate the reliability of the calibration curve, the air kerma exposed to nanoDots and the response of nanoDots to scattered X-rays, penetrating X-rays and/or the entrance surface dose due to the 83 kV X-rays were evaluated. In the same manner, 55 and 108 kV X-rays were irradiated. The responses of nanoDots exposed by each condition were in good agreement with the main calibration curve and the reliability of it is ±15% even in 3 μGy. We concluded that the dose of diagnostic X-rays can be evaluated by the response of nanoDot calibrated by the air Kerma of 83 kV X-rays. (author)
Availability note (English)
Available from https://doi.org/10.15669/pnst.6.22Additional details
Identifiers
- DOI
- 10.15669/pnst.6.22;
Publishing Information
- Journal Title
- Progress in Nuclear Science and Technology
- Journal Volume
- 6
- Journal Page Range
- p. 22-25
- ISSN
- 2185-4823
Conference
- Title
- 9. international symposium on radiation safety and detection technology
- Acronym
- ISORD-9
- Dates
- 10-14 Jul 2017
- Place
- Nagoya, Aichi (Japan)
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 50050073
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ABSORBED RADIATION DOSES; CALIBRATION; ENERGY DEPENDENCE; IONIZATION CHAMBERS; LIGHT EMITTING DIODES; LUMINESCENT DOSEMETERS; MEDICAL PERSONNEL; PATIENTS; PHANTOMS; PHOTOMULTIPLIERS; RADIATION PROTECTION; SCATTERING; TUMOR CELLS; X-RAY SOURCES; X-RAY SPECTRA
- Descriptors DEC
- ANIMAL CELLS; DOSEMETERS; DOSES; MEASURING INSTRUMENTS; MOCKUP; PERSONNEL; PHOTOTUBES; RADIATION DETECTORS; RADIATION DOSES; RADIATION SOURCES; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SPECTRA; STRUCTURAL MODELS
Optional Information
- Notes
- 22 refs., 2 figs.