Description

A comparative study of the n-γ discrimination done by the digital charge comparison and zero-crossing methods was carried out for a 130 mm in diameter and 130 mm high BC501A liquid scintillator coupled to a 130 mm diameter XP4512B photomultiplier. The high quality of the tested detector was reflected in a photoelectron yield of 2300±100 phe/MeV and excellent n-γ discrimination properties with energy discrimination thresholds corresponding to very low neutron (or electron) energies. The superiority of the Z/C method was demonstrated for the n-γ discrimination method alone, as well as, for the simultaneous separation by the pulse shape discrimination and the time-of-flight methods down to about 30 keV recoil electron energy. The digital charge comparison method fails for a large dynamic range of energy and its separation is weakly improved by time-of-flight method for low energies. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
360
Journal Issue
3
Journal Page Range
p. 584-592.
ISSN
0168-9002
CODEN