Published August 1995 | Version v1
Journal article

Silicon detector for high rate EXAFS applications

  • 1. Brookhaven National Lab., Upton, NY (United States)
  • 2. Politecnico di Milano (Italy). Dipt. di Elettronica e Informazione

Description

A multichannel silicon pad detector for EXAFS (Extended X-ray Absorption Fine Structure) applications has been designed and built. The X-ray spectroscopic measurements demonstrate that an adequate energy resolution of 230 eV FWHM (corresponding to 27 rms electrons in silicon) can be achieved reliably at -35 C. A resolution of 190 eV FWHM (corresponding to 22 rms electrons) has been obtained from individual pads at -35 C. At room temperature (25 C) an average energy resolution of 380 eV FWHM is achieved and a resolution of 350 eV FWHM (41 rms electrons) is the best performance. A simple cooling system constituted of Peltier cells is sufficient to reduce the reverse currents of the pads and their related shot noise contribution, in order to achieve resolutions better than 300 eV FWHM which is adequate for the EXAFS applications

Additional details

Additional titles

Augmented title (English)
Extended x-ray absorption fine structure

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
42
Journal Issue
4Pt1
Journal Page Range
p. 585-589.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
medical imaging conference.
Acronym
Nuclear science symposium
Dates
30 Oct - 5 Nov 1994.
Place
Norfolk, VA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
27015901
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
AMPLIFIERS; DESIGN; ENERGY RESOLUTION; PERFORMANCE; SENSITIVITY; SI SEMICONDUCTOR DETECTORS; USES; X-RAY SPECTROSCOPY
Descriptors DEC
EQUIPMENT; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTROSCOPY

Optional Information

Secondary number(s)
CONF-941061--.