Towards large area X- and gamma-ray imagers based on Controlled Drift Detectors

Description

The design solutions of a new generation of Controlled Drift Detectors (CDD) with larger area and flexible pixel size are discussed. The experimental results show that the active areas of few square centimetres and pixel sizes ranging from 50 to 180 μm can be safely designed keeping the readout time within few microseconds. Large-area CDD with excellent energy resolution and fast readout opens a variety of new biological, medical and industrial applications. Application examples to microsecond-scale time-resolved imaging of periodical processes, 2D/3D energy-weighted tomography and Compton telescope for γ-ray imaging with sub-millimeter resolution are proposed

Additional details

Identifiers

DOI
10.1016/j.nima.2003.11.048;
PII
S0168900203029024;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
518
Journal Issue
1-2
Journal Page Range
p. 426-428
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
Frontier detectors for frontier physics
Acronym
9. Pisa meting on advanced detectors
Dates
25-31 May 2003
Place
La Biodola, Isola d'Elba (Italy)