Published June 2003 | Version v1
Journal article

Development of RTS-200S 2D automatic scanning dose measuring water phantom for radiotherapy

  • 1. National Inst. for Radiological Protection and Nuclear Safety, China CDC, Beijing (China)

Description

Objective: To develop a 2D automatic scanning water phantom, being applicable for acceptance tests of radiotherapy equipment and quick measurements of clinical physical parameters for radiotherapy beams and suitable for use in radiotherapy departments and provincial monitoring institutions. Methods: Two chambers were used as a scanning detector and a beam monitor, respectively. Modern automatic techniques were adopted in the scanning system so as to realize rapid and accurate measurement for medical linear accelerator, X-knife and Co-60 therapy beams. The precision of scanning position was less than ±0.1 mm by means of computer-controlled stepper motors. Results: RTS-200S enabled user to acquire data of beam profiles, depth dose curve. RTP software program integrated data collection, analysis, output and storage. The measuring results were in agreement with those measured by Varian company. Conclusion: The RTS-200S water phantom and electronic control are excellent in both accuracy and reliability for measuring beam speciality of radiotherapy units

Additional details

Publishing Information

Journal Title
Chinese Journal of Radiological Medicine and Protection
Journal Volume
23
Journal Issue
3
Journal Page Range
p. 190-192
ISSN
0254-5098