Published 1985 | Version v1
Journal article

A consistent system for ionisation chamber calibration and measurements in air and in phantom

  • 1. Churchill Hospital, Oxford (UK). Dept. of Radiation Physics

Description

Measurements have been made to establish experimental support for the proposal that the dose to a point in an irradiated medium can be derived via in-air calibration of an ionisation chamber, with any material as a build-up cap. 60Co γ rays, 8 MV X rays and 18 MV X rays were used as the radiation sources, and build-up materials from Perspex to copper were studied. Experimental measurements have also shown that, for readings made in water, the presence or absence of a Perspex build-up cap does not alter the final measured dose. This finding is in agreement with theoretical predictions. (author)

Additional details

Publishing Information

Journal Title
Radiat. Prot. Dosim.
Journal Volume
12
Journal Issue
1
Series
Radiat. Prot. Dosim.
Journal Page Range
3-9
ISSN
0144-8420
CODEN
RPDOD