Published September 1990
| Version v1
Journal article
Gamma position-sensitive detector for positron-emission tomography
- 1. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Moscow (USSR). Inst. Atomnoj Ehnergii
Description
Use of hodoscopic photomultiplier for positron-emission tomography is suggested. Intrinsic spatial resolution of tomograph element is measured. For Bi4Ge3O12 scintillator it is 3.6 mm, for Ce: Gd2SiO5 - 1.7 mm. Structures with a single-layer and multilayer circular position of circular recording elements are suggested
Additional details
Additional titles
- Original title (Russian)
- Позиционно-чувствительный детектор гамма-квантов для позитронного эмиссионного томографа
Publishing Information
- Journal Title
- Pribory i Tekhnika Ehksperimenta
- Journal Issue
- no.5
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 23037065
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- EFFICIENCY; GAIN; HODOSCOPES; PHOTOCATHODES; PHOTOMULTIPLIERS; POSITION SENSITIVE DETECTORS; POSITRON COMPUTED TOMOGRAPHY; SOLID SCINTILLATION DETECTORS; SPATIAL RESOLUTION; TIME RESOLUTION
- Descriptors DEC
- AMPLIFICATION; CATHODES; COMPUTERIZED TOMOGRAPHY; ELECTRODES; EMISSION COMPUTED TOMOGRAPHY; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; TIMING PROPERTIES; TOMOGRAPHY