Published January 1983 | Version v1
Report Open

Measurement and simulation of the drift pulses and resolution in the micro-jet chamber

Description

We have tested a prototype of a micro-jet chamber, using both a nitrogen laser and a 10GeV electron beam. The achieved resolution in the particle beam was sigma = 18μm for a lmm impact parameter and 22μm when averaging over the entire beam profile. The experimental results were compared to a Monte Carlo program which simulates the pulse shapes and resolution in drift chambers of any geometry. The main emphasis in our simulation analysis was to study various strategies for drift chambers in order to achieve the best possible timing resolution

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE83012334.

Files

14788462.pdf

Files (297.4 kB)

Name Size Download all
md5:27569ab86690a29cb818502b5829eb99
297.4 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
16 p.
Report number
SLAC--PUB-3045

Conference

Title
Wire chamber conference.
Dates
15-18 Feb 1983.
Place
Vienna (Austria).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14788462
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
DRIFT CHAMBERS; ELECTRON BEAMS; GEV RANGE 01-10; MONTE CARLO METHOD; PARTICLE BEAMS; SIMULATION
Descriptors DEC
BEAMS; ENERGY RANGE; GEV RANGE; LEPTON BEAMS; MEASURING INSTRUMENTS; MULTIWIRE PROPORTIONAL CHAMBER; PROPORTIONAL COUNTERS; RADIATION DETECTORS

Optional Information

Secondary number(s)
CONF-830223--5.