Published 1991 | Version v1
Report Restricted

20 element HgI2 energy dispersive x-ray array detector system

  • 1. Xsirius, Inc., Marina Del Rey, CA (United States)
  • 2. Stanford Univ., CA (United States). Stanford Synchrotron Radiation Lab.
  • 3. EG and G Energy Measurements, Inc., Goleta, CA (United States)

Description

This paper describes recent progress in the development of HgI2 energy dispersive x-ray detector arrays and associated miniaturized processing electronics for synchrotron radiation research applications. The experimental results with a 20 element array detector were obtained under realistic synchrotron beam conditions at SSRL. An energy resolution of 250 eV (FWHM) at 5.9 keV (Mn-Ka) was achieved. Energy resolution and throughput measurements versus input count rate and energy of incoming radiation have been measured. Extended X-ray Absorption Fine Structure (EXAFS) spectra were taken from diluted samples simulating proteins with nickel

Availability note (English)

MF available from INIS under the Report Number; OSTI as DE92013356; NTIS; INIS; US Govt. Printing Office Dep.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
5 p.
Report number
EGG--10617-2135

Conference

Title
IEEE nuclear science symposium.
Dates
5-9 Nov 1991.
Place