Published November 26, 2013 | Version v1
Journal article

Validity of actigraphs uniaxial and triaxial accelerometers for assessment of physical activity in adults in laboratory conditions

  • 1. Department of Exercise Science, California Lutheran University, 60 W. Olsen Road, #3400, Thousand Oaks, CA 91360 (United States)
  • 2. Packet Photonics Inc, Santa Barbara, CA (United States)

Description

Few studies to date have directly compared the Actigraphs GT1M and the GT3X, it would be of tremendous value to know if these accelerometers give similar information about intensities of PA. Knowing if output is similar would have implications for cross-examination of studies. The purpose of the study was to assess the validity of the GT1M and the GT3X Actigraph accelerometers for the assessment of physical activity against oxygen consumption in laboratory conditions. Forty-two college-aged participants aged 18-25 years wore the GT1M and the GT3X on their right hip during treadmill exercise at three different speeds, slow walking 4.8 km.h-1, fast walking 6.4 km.h-1, and running 9.7 km.h-1). Oxygen consumption was measured minute-by minute using a metabolic system. Bland-Altman plots were used to assess agreement between activity counts from the GT3X and GT1M, and correlations were assessed the ability of the accelerometers to assess physical activity. Bias for 4.8 km.h-1 was 2814.4 cpm (limits 1211.3 to 4417.4), for 6.4 km.h-1 was 3713.6 cpm (limits 1573.2 to 5854.0), and for 9.7 km.h-1 was−3811.2 cpm (limits 842.1 to 6780.3). Correlations between counts per minute for the GT1M and the GT3X were significantly correlated with VO2 (r = 0.881, p < 0.001; r = 0.810, p < 0.001 respectively). The present study showed that both the GT1M and the GT3X accurately measure physical activity when compared to oxygen consumption

Availability note (English)

Available from http://dx.doi.org/10.1186/1756-6649-13-5; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4175111

Additional details

Publishing Information

Journal Title
BMC medical physics
Journal Volume
13
Journal Page Range
p. 5
ISSN
1756-6649

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46092605
Subject category
S42: ENGINEERING;
Descriptors DEI
ACCELEROMETERS; CHARGES; CORRELATIONS; EDUCATIONAL FACILITIES; EXERCISE; OXYGEN; VALIDATION; VELOCITY
Descriptors DEC
ELEMENTS; MEASURING INSTRUMENTS; NONMETALS; TESTING

Optional Information

Copyright
Copyright (c) 2013 Kelly et al.
Notes
PMCID: PMC4175111; PUBLISHER-ID: 1756-6649-13-5; PMID: 24279826; OAI: oai:pubmedcentral.nih.gov:4175111; licensee BioMed Central Ltd.