Published February 1985 | Version v1
Journal article

Quality control of radionuclide calibrators: Interlaboratory comparison of thallium-201

Creators

  • 1. Physikalisch-Technische Bundeanstalt, Braunschweig (Germany, F.R.). Lab. fuer Radioaktive Standards

Description

After 131I and sup(99m)Tc comparisons in 1982 and 1983, the Physikalisch-Technische Bundesanstalt (PTB) in cooperation with the firm Amersham-Buchler organized a further comparison in October 1984 in order to investigate the accuracy of activity measurements in nuclear medicine. 48 hospitals or institutes participated in this 201TI comparison. The wast majority of the reported results agreed with the PTB-value to within 10%, however in 5 cases deviations amounted to more than 20%. (orig.)

Abstract (German)

Nach der Durchfuehrung von 131J- und sup(99m)Tc-Ringversuchen in den Jahren 1982 und 1983 organisierte die Physikalisch-Technische Bundesanstalt (PTB) in Zusammenarbeit mit der Firma Amersham-Buchler im Oktober 1984 einen weiteren Ringversuch zur Ueberpruefung der Messsicherheit von Aktivimetern in der Nuklearmedizin. An diesem 201TI-Ringversuch nahmen 48 Krankenhaeuser, Kliniken oder Institute teil. Die gemeldeten Messwerte stimmen zum ueberwiegenden Teil (etwa 80%) mit den PTB-Werten innerhalb von 10% ueberein, allerdings betragen die Abweichungen in etwa 5 Faellen mehr als 20%. (orig.)

Additional details

Additional titles

Original title (German)
Qualitaetskontrolle von Aktivitaetsmessgeraeten:

Publishing Information

Journal Title
Nuc Compact
Journal Volume
16
Journal Issue
1
Series
CODEN: CNNMA.;Nuc Compact.
Journal Page Range
46-49
ISSN
0344-3752