The experimental investigation of coating's thickness measurement based on x-fluorescence
Creators
- 1. The School of Nuclear Science and Technology, the University of South China, Hengyang (China)
- 2. The School of Manufacturing Science and Engineering, Sichuan University, Chengdu (China)
Description
The control of coating's thickness is very important. If coating's thickness is not up to the standard, it will have a serious effect on the performance of the products. The test platform of coating's thickness measurement is built in the laboratory, Cu, Fe was used as the test sample, oil paint was used as coating in the experiment. The working curves are established after calibration using standard samples, taking advantage of the exponent attenuation law I=I0 e-μx, when ray is traversing matter; measuring coating's thickness on the base of Cu and Fe, under a condition that coating's thickness is from 0.01 g/cm2 to 0.06 g/cm2, the correlation coefficient is 0.992 and 0.995 respectively, indicate that the measuring method is feasible. (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 34
- Journal Issue
- 7
- Journal Page Range
- p. 866-868, 873
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 49079753
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ATTENUATION; CALIBRATION; CONTROL; ELECTRONS; FLUORESCENCE; MATTER; MEASURING METHODS; PERFORMANCE; THICKNESS
- Descriptors DEC
- DIMENSIONS; ELEMENTARY PARTICLES; EMISSION; FERMIONS; LEPTONS; LUMINESCENCE; PHOTON EMISSION
Optional Information
- Notes
- 7 figs., 2 tabs., 6 refs.