A phantom study to determine the optimal placement of eye dosemeters on interventional cardiology staff

  • 1. Department of Radiological Examination and Technology, Tohoku University Graduate School of Medicine, 2-1 Seiryo, Aoba-ku, Sendai 980-8575 (Japan)
  • 2. Disaster Medical Radiology, Division of Disaster Medical Science, International Research Institute of Disaster Science, Tohoku University, Aramaki Aza-Aoba 468-1, Aoba-ku, Sendai 980-0845 (Japan)
  • 3. Department of Radiology, Sendai Kousei Hospital, 4-5 Hirosemachi, Aoba-ku, Sendai 980-0873 (Japan)

Description

The International Commission on Radiological Protection has substantially reduced the recommended maximum annual eye lens dose for workers. Use of a dedicated eye dosemeter is one method for accurate dose monitoring. The main aim of this study was to yield recommendations for optimal placement of eye dosemeters to estimate the eye dose to interventional cardiology physicians and nurses. A phantom measurement was conducted to simulate typical interventional cardiology procedures. Considering eight X-ray tube angulations, the left side of the head position provide good estimates for physician, and the forehead position provide good estimates for nurse. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncz027

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
185
Journal Issue
4
Journal Page Range
p. 409-413
ISSN
0144-8420

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
51063008
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSEMETERS; DOSES; EYES; PHANTOMS; RADIATION PROTECTION; X-RAY TUBES
Descriptors DEC
BODY; ELECTRON TUBES; EQUIPMENT; FACE; HEAD; MEASURING INSTRUMENTS; MOCKUP; ORGANS; SENSE ORGANS; STRUCTURAL MODELS; X-RAY EQUIPMENT

Optional Information

Notes
23 refs.