Published October 1, 2019 | Version v1
Journal article

X-Tream dosimetry of synchrotron radiation with the PTW microDiamond

  • 1. Centre for Medical Radiation Physics, University of Wollongong, Northfields Avenue, Wollongong (Australia)

Description

In this study, we demonstrate that the X-Tream dosimetry system developed by the Centre for Medical Radiation Physics can be used with both commercial and non-commercial dosimeters. A custom adaptor was constructed to connect the PTW microdiamond to the X-Tream DAQ. This allowed for the measurements of continuous depth dose curves, broad beam and microbeam profiles. The X-Tream system enables real time measurement of the instantaneous dose rate of synchrotron generated radiation with the PTW microDiamond. Finally, measurements undertaken in this study demonstrate that the X-Tream system is able to mitigate the alignment issue that has previously hampered the uptake of the PTW microdiamond in clinical quality assurance for small field dosimetry.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/14/10/P10037

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
14
Journal Issue
10
Journal Page Range
p. P10037
ISSN
1748-0221