Published May 1983 | Version v1
Journal article

On macro-stereoangiography (optimum angle of deviation for x-ray tube)

  • 1. National Cardiovascular Center, Suita, Osaka (Japan)

Description

Because of the different inter-focal shifting distances, the macro-stereoangiographic X-ray tubes developed recently by various makers, naturally have differences in depth, for example, in case of observations and differ in steric effect according to the observating condition, as well. Studies made of these issues revealed that the angle of deviation for X-ray tube was optimum for 2 deg to 3 deg and the magnification (M) was proportional to 1/Msup(1.5). (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
39
Journal Issue
3
Series
Nippon Hoshasen Gijutsu Gakkai Zasshi.
Journal Page Range
330-332
ISSN
0369-4305