Published April 15, 1983 | Version v1
Journal article

SUPERLUMI: A high flux VUV spectroscopic device for luminescence measurement

  • 1. Hamburg Univ. (Germany, F.R.). 2. Inst. fuer Experimentalphysik
  • 2. Paris-11 Univ., 91 - Orsay (France). Lab. de Physique des Gaz et des Plasma

Description

A new experimental set up for time and spectrally resolved luminescence experiments on atoms, molecules and solids under state selective excitation with synchrotron radiation (SR) is described. A uniquely large spectral range of luminescence analysis from 50 to 1000 nm is covered by a combination of two monochromators. The central device is a specially designed VUV toroidal grating monochromator with an extremely large f-number 1:2.8 for a working range from 50 to 300 nm. Excellent agreement between calculated and measured performance of this instrument is found. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res.
Journal Volume
208
Journal Issue
1-3
Series
Nucl. Instrum. Methods Phys. Res.
Journal Page Range
835-839
ISSN
0029-554X

Conference

Title
International conference on X-Ray and VUV synchrotron radiation instrumentation.
Dates
9-13 Aug 1982.
Place
Hamburg (Germany, F.R.).

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
14796748
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
EXTREME ULTRAVIOLET RADIATION; LUMINESCENCE; MONOCHROMATORS; PHOTON BEAMS; SPECTROSCOPY; SYNCHROTRON RADIATION; TIME RESOLUTION
Descriptors DEC
BEAMS; BREMSSTRAHLUNG; ELECTROMAGNETIC RADIATION; RADIATIONS; RESOLUTION; TIMING PROPERTIES; ULTRAVIOLET RADIATION