Published January 2006 | Version v1
Journal article

Projection profile analysis for automated detection of abnormalities in chest radiographs

  • 1. Department of Medical Imaging, Kaohsiung Medical University Chung-Ho Memorial Hospital, Kaohsiung, Taiwan (China)
  • 2. Department of Computer Science and Engineering, National Sun Yat-Sen University, Kaohsiung (China)

Description

Abnormalities in chest images often present as abnormal opacity or abnormal asymmetry. We have developed a novel method for automated detection of abnormalities in chest radiographs by use of these features. Our method is based on an analysis of the projection profile obtained by projecting the pixels data of a frontal chest image on to the mediolateral axis. Two indices, lung opacity index and lung symmetry index, are computed from the projection profile. Lung opacity index and lung symmetry index are then combined to detect gross abnormalities in chest radiographs. The values of lung opacity index are found to be 0.38±0.05 and 0.37±0.06 for normal right and left lung, respectively. The values of lung symmetry index are found to be 0.018±0.014 for normal chest images. The discrimination for the combination of the two indices is evaluated by linear discriminant analysis and receiver operating characteristic (ROC) analysis. Area Az under the ROC curve with the combination of the two indices in the classification of normal and abnormal chest images is 0.963

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
33
Journal Issue
1
Journal Page Range
p. 118-123
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
37078535
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ASYMMETRY; CHEST; IMAGE PROCESSING; IMAGES; LUNGS; OPACITY; SENSITIVITY ANALYSIS
Descriptors DEC
BODY; OPTICAL PROPERTIES; ORGANS; PHYSICAL PROPERTIES; PROCESSING; RESPIRATORY SYSTEM

Optional Information

Notes