Comparing an accelerated 3D fast spin-echo sequence (CS-SPACE) for knee 3-T magnetic resonance imaging with traditional 3D fast spin-echo (SPACE) and routine 2D sequences
Creators
- 1. Northwestern University Feinberg School of Medicine, Department of Radiology, Chicago, IL (United States)
- 2. Marshfield Clinic, Department of Radiology, Marshfield, WI (United States)
- 3. Siemens Healthcare GmbH, Erlangen (Germany)
Description
To compare a faster, new, high-resolution accelerated 3D-fast-spin-echo (3D-FSE) acquisition sequence (CS-SPACE) to traditional 2D and high-resolution 3D sequences for knee 3-T magnetic resonance imaging (MRI). Twenty patients received knee MRIs that included routine 2D (T1, PD ± FS, T2-FS; 0.5 x 0.5 x 3 mm3; ∝10 min), traditional 3D FSE (SPACE-PD-FS; 0.5 x 0.5 x 0.5 mm3; ∝7.5 min), and accelerated 3D-FSE prototype (CS-SPACE-PD-FS; 0.5 x 0.5 x 0.5 mm3; ∝5 min) acquisitions on a 3-T MRI system (Siemens MAGNETOM Skyra). Three musculoskeletal radiologists (MSKRs) prospectively and independently reviewed the studies with graded surveys comparing image and diagnostic quality. Tissue-specific signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were also compared. MSKR-perceived diagnostic quality of cartilage was significantly higher for CS-SPACE than for SPACE and 2D sequences (p < 0.001). Assessment of diagnostic quality of menisci and synovial fluid was higher for CS-SPACE than for SPACE (p < 0.001). CS-SPACE was not significantly different from SPACE but had lower assessments than 2D sequences for evaluation of bones, ligaments, muscles, and fat (p ≤ 0.004). 3D sequences had higher spatial resolution, but lower overall assessed contrast (p < 0.001). Overall image quality from CS-SPACE was assessed as higher than SPACE (p = 0.007), but lower than 2D sequences (p < 0.001). Compared to SPACE, CS-SPACE had higher fluid SNR and CNR against all other tissues (all p < 0.001). The CS-SPACE prototype allows for faster isotropic acquisitions of knee MRIs over currently used protocols. High fluid-to-cartilage CNR and higher spatial resolution over routine 2D sequences may present a valuable role for CS-SPACE in the evaluation of cartilage and menisci. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00256-016-2490-8Additional details
Identifiers
Publishing Information
- Journal Title
- Skeletal Radiology
- Journal Volume
- 46
- Journal Issue
- 1
- Journal Page Range
- p. 7-15
- ISSN
- 0364-2348
- CODEN
- SKRADI
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 48019508
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BODY FLUIDS; BONE JOINTS; CARTILAGE; COMPARATIVE EVALUATIONS; DIAGNOSTIC USES; FEMUR; IMAGE PROCESSING; LIGAMENTS; MUSCLES; NMR IMAGING; SIGNAL-TO-NOISE RATIO; SPATIAL RESOLUTION; SPIN ECHO; TIBIA; TWO-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- ANIMAL TISSUES; BIOLOGICAL MATERIALS; BODY; CONNECTIVE TISSUE; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; EVALUATION; MATERIALS; ORGANS; PROCESSING; RESOLUTION; SKELETON; USES