Published June 2014 | Version v1
Journal article

Development of signal processing modules for double-sided silicon strip detector of gamma vertex imaging for proton beam dose verification

  • 1. Dept. of Nuclear Engineering, Hanyang University, Seoul (Korea, Republic of)
  • 2. Heavy-ion Clinical Research Division, Korea Institute of Radiological and Medical Sciences, Seoul (Korea, Republic of)

Description

Recently, a new imaging method, gamma vertex imaging (GVI), was proposed for the verification of in-vivo proton dose distribution. In GVI, the vertices of prompt gammas generated by proton induced nuclear interaction were determined by tracking the Compton-recoiled electrons. The GVI system is composed of a beryllium electron converter for converting gamma to electron, two double-sided silicon strip detectors (DSSDs) for the electron tracking, and a scintillation detector for the energy determination of the electron. In the present study, the modules of a charge sensitive preamplifier (CSP) and a shaping amplifier for the analog signal processing of DSSD were developed and the performances were evaluated by comparing the energy resolutions with those of the commercial products. Based on the results, it was confirmed that the energy resolution of the developed CSP module was a little lower than that of the CR-113 (Cremat, Inc., MA), and the resolution of the shaping amplifier was similar to that of the CR-200 (Cremat, Inc., MA). The value of Vrms representing the magnitude of noise of the developed system was estimated as 6.48 keV and it was confirmed that the trajectory of the electron can be measured by the developed system considering the minimum energy deposition ( > ⁓51 keV) of Compton-recoiled electron in 145-μm-thick DSSD.

Additional details

Publishing Information

Journal Title
Journal of Radiation Protection
Journal Volume
39
Journal Issue
2
Series
15 refs, 9 figs, 1 tab
Journal Page Range
p. 81-89
ISSN
0253-4231