The energy dependence for each of five models of low-energy direct-reading pocket dosimeters was characterized through an examination of the exposure correction factors determined over the diagnostic x-ray range (0.4--4.1-mm Al HVL) for many samples of each model. The results of an analysis of variance performed by model on the correction factors are reported as mean correction factors for the x-ray beams evaluated, the relative standard error being less than 0.8% in all cases. Energy-dependence curves for the dosimeter models are given, and their use is described

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
4
Journal Issue
6
Series
Med. Phys.
Journal Page Range
503-504
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
9374346
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CORRECTIONS; DOSEMETERS; ENERGY DEPENDENCE; RESPONSE FUNCTIONS; X-RAY DOSIMETRY
Descriptors DEC
DOSIMETRY; FUNCTIONS; MEASURING INSTRUMENTS