Published May 2010 | Version v1
Book

Flat detector versus image intensifier in cardiology. Why are patient doses initially higher?

  • 1. San Carlos University Hospital and Complutense University, 28040 Madrid (Spain)
  • 2. Az. Ospedale S. Maria della Misericordia, 33100 Udine (Italy)

Description

Our purpose in this paper is to study the differences in patient doses imparted using dynamic flat detectors (FD) as opposed to using image intensifier (II) technology for interventional cardiology within two different European Centres, and then analyse the reasons for such differences. Several x-ray systems from two different manufacturers equipped with FDs ands IIs were used over the last 4 years in two separate interventional cardiology services both subject to common quality assurance programmes agreed during the European SENTINEL Coordination Action. The same team of cardiologists worked simultaneously or in sequence in laboratories equipped with FD and II technologies. No differences were found regarding specific patient characteristics in the different catheterisation laboratories. Kerma area product (KAP) values were measured and analysed for large samples of procedures. In centre M, median KAP values were consistently and significantly higher with FD than with II, with percentages of 55-113%. In centre U, the KAP increase was 7-29% comparing FD with II. Without a dedicated optimisation programme, initial patient dose values can result significantly higher with FD technology in comparison with II. This might be related to the use of non optimum initial setting protocols of the x-ray systems (dose per frame, pulses/s, filtration etc), bigger radiation filed sizes for FD and the scarce use of collimation and wedge filters in the new systems equipped with FDs. Preventing this problem and improving the management of patient doses warrant further studies. (author)

Part of:
Proceedings of IRPA12: 12. Congress of the International Radiation Protection Association: Strengthening Radiation Protection Worldwide - Highlights, Global Perspective and Future Trends

Additional details

Publishing Information

Publisher
IAEA
Imprint Place
Vienna (Austria)
ISBN
978-92-0-105410-4
Imprint Title
Proceedings of IRPA12: 12. Congress of the International Radiation Protection Association: Strengthening Radiation Protection Worldwide - Highlights, Global Perspective and Future Trends
Imprint Pagination
vp.
Series
Proceedings Series
Journal Page Range
6 p.
ISSN
0074-1884

Conference

Title
12. Congress of the International Radiation Protection Association: Strengthening Radiation Protection Worldwide - Highlights, Global Perspective and Future Trends
Acronym
IRPA12
Dates
19-24 Oct 2008
Place
Buenos Aires (Argentina)

INIS

Country of Publication
Austria
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42077978
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
IMAGE INTENSIFIERS; KERMA; PATIENTS; QUALITY ASSURANCE; RADIATION DOSES; X-RAY EQUIPMENT
Descriptors DEC
DOSES; EQUIPMENT

Optional Information

Contract/Grant/Project number
Grant FIS2006-08186
Notes
10 refs, 4 figs, 1 tab Imprint:This CD-ROM is attached to the printed publication
Funding organization
Ministry of Science and Innovation (Spain)
Secondary number(s)
STI/PUB--1460 (Companion CD)