Published January 2005 | Version v1
Journal article

Electronic compensation using multileaf collimation for involved field radiation to the neck and mediastinum in non-Hodgkin's lymphoma and Hodgkin's lymphoma

  • 1. Department of Medical Physics, Tom Baker Cancer Centre, Calgary, Alberta (Canada)
  • 2. Department of Oncology, University of Calgary, Calgary, Alberta (Canada)
  • 3. Department of Radiation Oncology, Tom Baker Cancer Centre, Calgary, Alberta (Canada)
  • 4. Department of Medical Physics, Tom Baker Cancer Centre, Calgary, Alberta (Canada) and Department of Oncology, University of Calgary, Calgary, Alberta (Canada)

Description

An efficient procedure is required for the preparation, planning, and delivery of radiation therapy for involved field radiation to the neck and mediastinum. This technique must reduce tissue complications while maintaining dose uniformity. An elegant intensity-modulated radiation therapy (IMRT) treatment that is forward planned has been developed. Both static fields and static subfields shaped by multileaf collimators (MLCs) and asymmetric jaws are used. Patients receiving involved field radiation to the neck and mediastinum are planned in 3 dimensions (3D), where 3D dose compensation is provided using subfields consisting of MLC or asymmetric jaws instead of physical compensators or wedges. Forward planning is performed, usually generating 2 pairs of parallel-opposed fields, with at least 1 of them consisting of subfields to eliminate elevated dose regions. Efficiency in the preparation, planning, and delivery of treatment has been achieved for more than 10 patients. Verification of treatment setup, target anatomy, and MLC configuration is quick when using an electronic portal imaging device. Thermoluminescent dosimeters (TLDs) have verified point-dose uniformity noticeably to ± 5%. An efficient technique using forward planning for simple IMRT consisting of static MLC and asymmetric jaws has been developed

Additional details

Identifiers

DOI
10.1016/j.meddos.2004.11.001;
PII
S0958-3947(04)00164-5;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
30
Journal Issue
2
Journal Page Range
p. 59-64
ISSN
0958-3947

Optional Information

Copyright
Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.