Published February 1992 | Version v1
Journal article

A physical image quality evaluation of a digital spot fluorography system

  • 1. Leeds General Infirmary (United Kingdom). Dept. of Nuclear Medicine

Description

A physical evaluation of the imaging performance of the first digital spot fluorography system installed in the UK is presented. Measurements have been made of dynamic range, unsharpness (MTF), noise (Wiener spectrum) and threshold contrast detail detectability (TCDD) and the results compared with those produced by conventional spot-film radiography systems. Noise-equivalent parameters are derived from the objective measurements and the results related to the findings of the TCDD tests. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
37
Journal Issue
2
Series
Phys. Med. Biol.
Journal Page Range
325-342
ISSN
0031-9155
CODEN
PHMBA

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
23039498
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; CAMERA TUBES; COMPARATIVE EVALUATIONS; DIGITAL SYSTEMS; FLUOROSCOPY; IMAGE INTENSIFIERS; IMAGE PROCESSING; NOISE; RADIATION DOSES; SPATIAL RESOLUTION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; EVALUATION; IMAGE TUBES; MEDICINE; RESOLUTION