Physical characterizations for an integrated 160-leaf multi-leaf collimator with a new concept design
- 1. Kumamoto Univ. Hospital, Kumamoto, Kumamoto (Japan)
- 2. Kumamoto Univ., Faculty of Life Sciences, Kumamoto, Kumamoto (Japan)
Description
In this article, we present a physical characterization of the agilityTM (Elekta). agilityTM is composed of 160 interdigitating multileaf collimators (MLCs) with a width of 5 mm at the isocenter. The physical characterizations that include leaf position accuracy, leakage, field penumbra and the tongue-and-groove (T and G) effect were evaluated using well-commissioned 4, 6 and 10-MV photon beams. The leaf position accuracy was within 0.5 mm for all gantry angles and each MLC. The leakage was 0.44% on average and reached 0.47% at 10 MV: remarkably low due to a new design with tilted leaves. However, the T and G effect occurred due to tilt. It was approximately 20.8% on average and reached 22.3% at 6 MV. The penumbra width increased up to 8.5 mm at a field size of 20 x 20 cm at 4 MV. High position designed MLCs create a wider penumbra but show lower leakage and large head clearance. Head clearance is an important factor in stereotactic radiotherapy with multiple non-coplanar beams. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 69
- Journal Issue
- 7
- Journal Page Range
- p. 778-783
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 45019969
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COLLIMATORS; IONIZATION CHAMBERS; LEAKS; PHANTOMS; PHOTON BEAMS; RADIATION DOSES; RADIATION TRANSPORT; X RADIATION
- Descriptors DEC
- BEAMS; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MOCKUP; RADIATION DETECTORS; RADIATIONS; STRUCTURAL MODELS