Performance verification of the Suzaku X-ray Spectrometer in the flight configuration
Creators
- 1. RIKEN, 2-1 Hirosawa, Wako, Saitama 351-0198 (Japan)
- 2. NASA GSFC, Greenbelt, Maryland 20771 (United States)
- 3. LLNL, 7000 East Avenue, L-260, Livermore, CA 94550 (United States)
- 4. ISAS/JAXA, 3-1-1 Yoshinodai, Sagamihara, Kanagawa 229-8510 (Japan)
- 5. Tokyo Metropolitan University, 1-1 Minami-osawa, Hachioji, Tokyo 192-0397 (Japan)
- 6. University of Wisconsin, 1150 University Ave. Madison, WI 53706 (United States)
- 7. Miyazaki University, 1-1 Gakuen Kibanadai-nish, Miyazaki 889-2192 (Japan)
Description
The X-ray Spectrometer (XRS) is a high resolution, non-dispersive cryogenic detector on board the X-ray satellite, Suzaku (Astro-E2), which was successfully launched on July 10, 2005. The XRS achieves an energy resolution of 6eV at 6keV (FWHM) and covers a broad energy range of ∼0.07-10keV. The XRS will enable powerful plasma diagnostics of a variety of astrophysical objects such as the dynamics of gas in clusters of galaxies. The XRS was integrated to the spacecraft in September 2004, and underwent a series of spacecraft tests until April 2005. We describe results of the XRS performance verification in the spacecraft configuration. First, the noise level was extremely low on the spacecraft, and most of the pixels achieved an energy resolution of 5-6eV at 5.9keV. Microphonic noise from the mechanical cooler was one of the concerns, but this was not a problem, when the dewar was integrated to the spacecraft and filled with solid neon. To attain the best energy resolution, however, correction of gain drift is mandatory. The XRS has a dedicated calibration pixel for that purpose, and drift correction using the calibration pixel is very effective when the gain variation is due to changes in the ∼60mK heat sink temperature. On the other hand, the calibration pixel and the other pixels do not respond in the same way to variations of the helium and neon bath temperatures, and this effect requires further study
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2005.12.100;
- PII
- S0168-9002(05)02531-3;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 559
- Journal Issue
- 2
- Journal Page Range
- p. 614-616
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 11. international workshop on low temperature detectors
- Acronym
- LTD-11
- Dates
- 31 Jul - 5 Aug 2005
- Place
- Tokyo (Japan)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37094058
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CALORIMETERS; CONFIGURATION; ENERGY RESOLUTION; EV RANGE 10-100; EV RANGE 100-1000; GALAXY CLUSTERS; HELIUM; KEV RANGE 01-10; NEON; PERFORMANCE; SATELLITES; VERIFICATION; X RADIATION; X-RAY SPECTROMETERS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; ELEMENTS; ENERGY RANGE; EV RANGE; FLUIDS; GASES; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; NONMETALS; RADIATIONS; RARE GASES; RESOLUTION; SPECTROMETERS
Optional Information
- Copyright
- Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.