Improving mass spectrometer sensitivity using a high-pressure electrodynamic ion funnel interface
Creators
- 1. Pacific Northwest National Laboratory, Biological Sciences Division (United States)
Description
We report on a new electrodynamic ion funnel that operates at a pressure of 30 torr with no loss of ion transmission. The enhanced performance compared with previous ion funnel designs optimized for pressures of <5 torr was achieved by reducing the ion funnel capacitance and increasing the RF drive frequency (1.7 MHz) and amplitude (100–170 V peak-to-peak). No degradation of ion transmission was observed for pressures from 2 to 30 torr. The ability to operate at higher pressure enabled a new tandem ion funnel mass spectrometer interface design that can accommodate a greater gas load (e.g., from an ESI source). When combined with a multicapillary inlet, the interface provided more efficient introduction of ions, resulting in a significant enhancement in mass spectrometer sensitivity and detection limits.
Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of the American Society for Mass Spectrometry
- Journal Volume
- 17
- Journal Issue
- 9
- Journal Page Range
- p. 1299-1305
- ISSN
- 1044-0305
- CODEN
- JAMSEF
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 50017234
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; ELECTRODYNAMICS; IONS; MASS SPECTROMETERS; SENSITIVITY
- Descriptors DEC
- CHARGED PARTICLES; EVALUATION; MEASURING INSTRUMENTS; SPECTROMETERS
Optional Information
- Copyright
- Copyright (c) 2006 American Society for Mass Spectrometry