Published September 2006 | Version v1
Journal article

Improving mass spectrometer sensitivity using a high-pressure electrodynamic ion funnel interface

  • 1. Pacific Northwest National Laboratory, Biological Sciences Division (United States)

Description

We report on a new electrodynamic ion funnel that operates at a pressure of 30 torr with no loss of ion transmission. The enhanced performance compared with previous ion funnel designs optimized for pressures of <5 torr was achieved by reducing the ion funnel capacitance and increasing the RF drive frequency (1.7 MHz) and amplitude (100–170 V peak-to-peak). No degradation of ion transmission was observed for pressures from 2 to 30 torr. The ability to operate at higher pressure enabled a new tandem ion funnel mass spectrometer interface design that can accommodate a greater gas load (e.g., from an ESI source). When combined with a multicapillary inlet, the interface provided more efficient introduction of ions, resulting in a significant enhancement in mass spectrometer sensitivity and detection limits.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of the American Society for Mass Spectrometry
Journal Volume
17
Journal Issue
9
Journal Page Range
p. 1299-1305
ISSN
1044-0305
CODEN
JAMSEF

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
50017234
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COMPARATIVE EVALUATIONS; ELECTRODYNAMICS; IONS; MASS SPECTROMETERS; SENSITIVITY
Descriptors DEC
CHARGED PARTICLES; EVALUATION; MEASURING INSTRUMENTS; SPECTROMETERS

Optional Information

Copyright
Copyright (c) 2006 American Society for Mass Spectrometry