Published March 1998 | Version v1
Journal article

An investigation into lung correction for tangential breast irradiation

  • 1. Townsville General Hospital, QLD (Australia). Medical Physics Department
  • 2. QRI Mater Centre, Raymond Terrace, South Brisbane, QLD (Australia)

Description

A solid water breast phantom with a lung equivalent material was constructed to investigate interface effects of 6 MV tangential photon fields. The phantom was built so that Thermo-luminescence Dosimeters (TLD) could be used for comparison to a pencil beam and Equivalent Tissue-Air-Ratio (ETAR) algorithms. The investigation highlighted the limitations in both of these algorithms under clinically anatomical conditions. At the lung/tissue interface both algorithms showed errors between 3-12% depending on the position of the TLD. In tissue greater than 1 cm from the interface region the errors on average were within + 5%. Dose comparisons in lung tissue showed similar results to the interface region. These results indicate that normalisation of Reference Dose in the interface region for tangential breast treatments cause over or under estimates of dose. Copyright (1998) Australasian Physical and Engineering Sciences in Medicine

Additional details

Publishing Information

Journal Title
Australasian Physical and Engineering Sciences in Medicine
Journal Volume
21
Journal Issue
1
Journal Page Range
p. 24-28
ISSN
0158-9938
CODEN
AUPMDI

INIS

Country of Publication
Australia
Country of Input or Organization
Australia
INIS RN
29066079
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
ALGORITHMS; CALIBRATION; CHEST; DOSE RATES; ERRORS; EXPERIMENTAL DATA; LUNGS; PHANTOMS; RADIATION DOSES; RADIOTHERAPY; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
BODY; BODY AREAS; DATA; DOSIMETRY; INFORMATION; MEDICINE; MOCKUP; NUMERICAL DATA; ORGANS; RESPIRATORY SYSTEM; STRUCTURAL MODELS; THERAPY

Optional Information

Notes
16 refs., 3 figs.