Published March 15, 1990 | Version v1
Journal article

Manufacturing avalanche diodes for detection of low energy gamma- and X-rays

  • 1. Sofia Univ. (Bulgaria). Fizicheski Fakultet
  • 2. Research Inst. TIEPE, Sofia (Bulgaria)
  • 3. Bylgarska Akademiya na Naukite, Sofia. Inst. po Yadrena Fizika i Yadrena Energetika

Description

Three types of silicon avalanche detectors for low energy gamma- and X-ray counting were manufactured using planar and planar-epitaxial technology. The detectors were tested at room temperature with 57Co and 241Am. Besides the characteristic lines of the sources additional peaks were observed due to X-rays of elements included in the matrices and holders of the sources. For 6.4 keV the energy resolution is approximately 10%. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
288
Journal Issue
2/3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
460-462
ISSN
0168-9002
CODEN