Published March 24, 2008 | Version v1
Journal article

Optimisation and critical evaluation of a collision cell technology ICP-MS system for the determination of arsenic in foodstuffs of animal origin

  • 1. Unite des Contaminants Inorganiques et Mineraux de l'Environnement, AFSSA-LERQAP 23, Avenue du G. de Gaulle, F-94706 Maisons-Alfort Cedex (France)

Description

The determination of arsenic (75As) was studied using an ICP-MS equipped with collision cell technology (CCT). Different mixtures of gases (He and H2) were tested using HCl conditions and a He flow rate of 4 mL min-1 was found to be suitable for the removal of the poly-atomic spectral interference [40Ar35Cl]+. Trueness of the optimised method has been evaluated in both standard and CCT modes on six certified reference materials in foodstuffs of animal origin and on three external proficiency testing schemes (FAPAS). The results obtained generally coincided with the certified values, except for CCT mode in some categories of samples (meat, mussels and milk powder), for which a positive bias on results was observed due to the formation of poly-atomic interferences within the collision cell. The main interferences were studied and their contributions estimated. [58Fe16O1H]+ and [74Ge1H]+ were the most significant interferences formed in the cell. Finally, different parameters (e.g. hexapole and quadrupole bias voltage, nebuliser gas flow) were optimised to try to attenuate these interferences

Availability note (English)

Available from http://dx.doi.org/10.1016/j.aca.2008.02.018

Additional details

Identifiers

DOI
10.1016/j.aca.2008.02.018;
PII
S0003-2670(08)00340-1;

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
611
Journal Issue
2
Journal Page Range
p. 134-142
ISSN
0003-2670
CODEN
ACACAM