Published November 1, 2018 | Version v1
Journal article

Organ at risk dose verification in nasopharyngeal cancer VMAT planning with collimator comparison between single arc and double arc

  • 1. Physics Department, University of Indonesia, Depok (Indonesia)
  • 2. Radiotherapy Department, Siloam Hospitals Simatupang, Jakarta (Indonesia)

Description

Volumetric modulated arc therapy (VMAT) is an irradiation technique that enables the improvement of the flexibility of dose delivery from a full range of angles with continuous modulation of beam aperture and dose rate. We compared the homogeneity index (HI), conformity index (CI), dose volume histogram (DVH) and dose differentiation of single arc and double arc VMAT delivered for nasopharyngeal cancer (NPC). Planning for gantry rotation angles of 360° counter clockwise and clockwise (double arc) with collimator (0° and 10°) and 360° clockwise (single arc) were created for ten nasopharyngeal cancer cases applied in RANDO phantom with the Eclipse treatment planning system and delivered using Trilogy linear accelerator. The verification was performed by using gafchromic film EBT3 and placed on the RANDO phantom according to the location of the organs at risk. In this study, we evaluated right parotid, left parotid, right eye, left eye and trachea. The measurement of mean dose using gafchromic film and TPS shows that the resulting values were close to TPS results with the difference being −1.51%. From the result of dose measurement using gafchromic film based on the data from RANDO CT phantom use, it can be concluded that double arc VMAT technique with collimator can reduce the dose in the organ at risk. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1757-899X/432/1/012024

Additional details

Publishing Information

Journal Title
IOP Conference Series. Materials Science and Engineering (Online)
Journal Volume
432
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
1757-899X

Conference

Title
1. Materials Research Society Indonesia Conference and Congress
Dates
8-12 Oct 2017
Place
Yogyakarta (Indonesia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52100439
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COLLIMATORS; COMPUTERIZED TOMOGRAPHY; DOSE RATES; IRRADIATION; LINEAR ACCELERATORS; MODULATION; NEOPLASMS; RADIATION HAZARDS; RADIOTHERAPY; THERAPEUTIC DOSES; THIN FILMS; VERIFICATION
Descriptors DEC
ACCELERATORS; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; FILMS; HAZARDS; HEALTH HAZARDS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY; TOMOGRAPHY