Published April 1985
| Version v1
Journal article
Beta dosimetry with newly developed graphite mixed TL detectors
Description
The characteristics of thermoluminescent dosemeters prepared from graphite mixed with sintered MgB4O7:Dy are presented for β ray dosimetry. These dosemeters are attractive for β ray dosimetry since the material combines low transparency with near tissue equivalence and high sensitivity, thus enabling skin dose to be assessed from low energy β emitters in accordance with the ICRP recommendations. Results from practical field experiments are also presented. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 30
- Journal Issue
- 4
- Series
- Phys. Med. Biol.
- Journal Page Range
- 323-329
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16058527
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BETA DOSIMETRY; BORATES; DYSPROSIUM; GRAPHITE; MAGNESIUM COMPOUNDS; RADIATION DOSES; SKIN; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- ALKALINE EARTH METAL COMPOUNDS; BODY; BORON COMPOUNDS; CARBON; DOSEMETERS; DOSIMETRY; ELEMENTS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; METALS; NONMETALS; ORGANS; OXYGEN COMPOUNDS; RARE EARTHS