Published April 1985 | Version v1
Journal article

Beta dosimetry with newly developed graphite mixed TL detectors

Creators

  • 1. Institut za Nuklearne Nauke Boris Kidric, Belgrade (Yugoslavia)

Description

The characteristics of thermoluminescent dosemeters prepared from graphite mixed with sintered MgB4O7:Dy are presented for β ray dosimetry. These dosemeters are attractive for β ray dosimetry since the material combines low transparency with near tissue equivalence and high sensitivity, thus enabling skin dose to be assessed from low energy β emitters in accordance with the ICRP recommendations. Results from practical field experiments are also presented. (author)

Additional details

Publishing Information

Journal Title
Phys. Med. Biol.
Journal Volume
30
Journal Issue
4
Series
Phys. Med. Biol.
Journal Page Range
323-329