X-ray beams characterization by the attenuation method for the personal dosemeters calibration

  • 1. Direccion General de Seguridad Radiologica, Instituto Peruano de Energia Nuclear, Lima (Peru)
  • 2. Facultad de Ciencias Fisicas, Universidad Nacional Mayor de San Marcos, Lima (Peru)

Description

The choice of the attenuation method to characterize the x-ray beams used in the calibration of personal dosemeters at the Calibration and Dosimetry Laboratory of the Peruvian Institute of Nuclear Energy (IPEN). Voltages applied to the X-ray tube and the addition of filters have been made according to the publication ANSI No. 13. Half value layer, homogeneity coefficient and mean energy were determined from attenuation data. Average relative differences of 3, 40 and 4% respectively were found with respect to values published in ANSI No. 13. Also spectral distributions and their resolutions were determined. relative differences of 6 and 19% respectively were found in comparison with values calculated by the Monte Carlo Method

Additional details

Additional titles

Original title (Spanish)
Caracterizacion de haces de rayos-x por el metodo de atenuacion para la calibracion de dosimetros personales