Object characterization simulator for estimating compressed breast during mammography
Creators
- 1. Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil). Post-graduation in Sciences and Technology of Radiations, Minerals and Materials
- 2. Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)
- 3. Departamento de Radiologia. Universidad Complutense de Madrid (Spain)
Description
The measurement of the thickness of a compressed breast during the mammography test is necessary in order to calculate the glandular dose in mammography procedures, in an analysis of risk/benefit, given that the target organ in these procedures is highly sensitive to ionising radiation. However, mammography is a test of utmost importance in diagnosis. In theory, it may be possible to calculate the thickness of the compressed breast through the measurements of the focus object distance by using projections of radio opaque objects fixed to the compression tray. The facilities of the Laboratory of Applied Radioprotection to Mammography - LARAM were used for this study, as well as breast simulators with well defined thickness, in the assembly of the techniques for the measurement of the thickness of the compressed breast. The results showed that it is possible to determine this thickness through calculations and simulators through this method which is susceptible to be adequate to the dosimetry. (author)
Files
43050761.pdf
Files
(172.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:fe01fbb227abfcd56c604e8e7a2477cf
|
172.0 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- INIS-BR--11020
Conference
- Title
- International nuclear atlantic conference. Nuclear energy: new jobs for a better life; 17. ENFIR: Meeting on nuclear reactor physics and thermal hydraulics; 10. ENAN: Meeting on nuclear applications; 2. ENIN: Meeting on nuclear industry
- Acronym
- INAC 2011
- Dates
- 24-28 Oct 2011
- Place
- Belo Horizonte, MG (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 43050761
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CARCINOGENS; CHEST; DIAGNOSTIC TECHNIQUES; GAMMA DETECTION; GAMMA DOSIMETRY; MAMMARY GLANDS; NEOPLASMS; NUCLEAR MEDICINE; RISK ASSESSMENT; SAFETY ANALYSIS; X-RAY DETECTION; X-RAY RADIOGRAPHY
- Descriptors DEC
- BODY; DETECTION; DISEASES; DOSIMETRY; GLANDS; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; MEDICINE; NONDESTRUCTIVE TESTING; ORGANS; RADIATION DETECTION; TESTING